LEEC CO2 Incubators Touch T50 - T50S

LEEC mini Culture Safe Touch CO2 incubators are the perfect solution for laboratory users who want a high quality, fully featured premium bench top mini CO2 incubator, in a small package.

Main features:

  • World’s first 200°C sterilisation (T50S)
  • 50 Litre Capacity
  • Touch screen controller
  • Deep drawn, seamless electro-polished chamber
  • Full data logging (text & graphical)
LEEC mini Culture Safe Touch CO2 incubators have a fully sealed, one-piece, deep drawn chamber, which is manufactured from 304 grade stainless steel. The outer cabinet is manufactured from stainless steel, and has an anti-bacterial polyester powder coat paint finish as standard. This gives the CO2 incubator exceptional resistance to corrosion and contamination. An inner glass viewing door closes positively onto an inert VOC free gasket (no volatile organic compounds) and allows a clear view into the chamber without disturbing the internal atmosphere. The outer door has a magnetic gasket providing a good seal. A 24mm rear access port allows temperature monitoring probes to be inserted safely into the CO2 incubator chamber.
The multi-function, PID colour touch screen controller regulates the temperature, CO2, alarm functions (including low gas and power failure alarms) and provides real time data logging in graphical and text format. The controller also allows data to be downloaded onto a removable SD card. An independent safety cut out disconnects the heating system should an over temperature condition occur. The controller can be password protected, stopping any unauthorised adjustments to the control systems by other laboratory staff.
TECHNICAL SPECIFICATIONS: LEEC mini Culture Safe Touch 50 / Touch 50S
TEMPERATURE MANAGEMENT CO2 GAS MANAGEMENT DIMENSIONS (w x h x d)
Range 5°C above ambient to +60°C Range 0% to 20% Chamber (mm) 402 x 402 x 300
Control 0.1°C Control ±0.1% Chamber (inches) 15.8 x 15.8 x 11.8
Stability  ± 0.10°C @ 37°C Stability ±0.2% External (mm) 534 x 717 x 460
Uniformity  ± 0.25°C @ 37°C Uniformity ±0.2% External (inches)  21.0 x 28.2 x 18.1
ELECTRICAL REQUIREMENTS CO2 Sensor Infra-Red Shipping Container (mm) 800 x 960 x 600
Voltage

230V AC 50/60Hz

(120V models available)

CO2 inlet pressure min 5 psi max 10 psi Shipping Container (inches) 31.5 x 37.8 x 23.6
Power  350w WEIGHT
Power, High Temp. Option  950w (T50S) CO2 inlet pressure min 0.3 bar max 0.7 bar Net

38 Kg (T50)

40 Kg (T50S)

Consumption to 37°C  0.072 kwh HUMIDITY   Shipping

50 Kg (T50)

52 Kg (T50S)

RH (@37°C + 5% upto 90%

Low Oxygen / Multi Gas incubator option.

LEEC mini Culture Safe CO2 incubator model Touch 50S is available with the OXY1-19 option. When factory fitted this allows low oxygen work inside the chamber down to 1% O2.

TECHNICAL SPECIFICATIONS:    LEEC mini Culture Safe Touch 50S with OXY1-19 option
TEMPERATURE CO2 O2 DIMENSIONS  (w x h x d)
Range 5°C above ambient to +60°C Range 0% to 20% Range 1% to 19% Chamber (mm) 402 x 402 x 300
Control 0.1°C Control ±0.1% Control ±0.1% Chamber (inches) 15.8 x 15.8 x 11.8
Stability ± 0.1°C @ 37°C Stability  ±0.2% Stability ±0.25% External (mm) 534 x 717 x 460
Uniformity ± 0.25°C @ 37°C Uniformity ±0.2% Unifomity ±0.25% External (inches) 21.0 x 28.2 x 18.1
ELECTRICAL CO2 Sensor Infra-Red O2 sensor Electrochemical Shipping container (mm) 800 x 960 x 600
Voltage

230V AC 50/60 Hz

(120V version available)

CO2 inlet pressure  min 5 psimax 10 psi O2 inlet pressure

min 5 psi

max 10 psi

Shipping container (inches) 31.5 x 37.8 x 26.6
Power (normal) 350w CO2 inlet pressure min 0.3 barmax 0.7 bar O2 inlet pressure

min 0.3 bar

max 0.7 bar

WEIGHT
Power (during sterilisation) 950w (T50S) Net 40 Kg (T50S)
Consumption at 37°C HUMIDITY  
0.072 kwh RH (@37°C + 5%) up to 90% Shipping 52 Kg (T50S)