Technical data

Specification Details
Working-temperature range +7 above ambient temperature up to +90 °C
Setting temperature range +18 to +90 °C
Setting accuracy temperature 0.1 °C
Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity Control Active microprocessor control for humidifying and dehumidifying (20 – 95 % rh), incl. digital indication and auto-diagnostic system ensures even more rapid reaching of set humidity and very short recovery times. Humidity supply with water (only for demineralised water with a conductivity of 5 to 10 μS/cm and a pH value between 5 and 7; from an external tank) by a self-priming pump; integral bacteria block by generating hot steam, dehumidifying via sterile filter
Setting range active humidity control 20 to 95 % rh and rh-Off
Setting accuracy humidity 0.5 % rh
ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting German, English, Spanish, French, Polish, Czech, Hungarian
adjustable parameters temperature (Celsius or Fahrenheit), relative humidity, programme time, time zones, summertime/wintertime
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable values each, temperature and humidity
Interface Ethernet LAN, USB
Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
AutoSAFETY additionally integrated over- and undertemperature monitor “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic system integral fault diagnostics for temperature and humidity control
Alarm visual and acoustic
6 sides large-area multi-function heating system on four sides with additional door and back heating to avoid condensation
Works calibration certificate Works calibration certificate (measuring point chamber centre) at 60 °C with 75 % rh
Door Additional heated inner glass door
Door fully insulated stainless steel door with 2-point locking (compression door lock)
Entry port Entry port (silicone), 40 mm clear diameter, for introducing connections, moisture tight, can be closed by silicone stopper, at the back, centre left
Internals 1 perforated stainless steel shelf/shelves
Standard accessories Water tank including connection hose
Stainless steel interior Dimensions w(A) x h(B) x d(C): 400 x 425 x 330 mm (d less 35 mm for fan)
Interior W.-St. 1.4301 (ASTM 304)
Volume 56 l
Max. number of internals 5
Max. loading of chamber 75 kg
Max. loading per internal 15 kg
Textured stainless steel casing Dimensions w(D) x h(E) x d(F): 559 x 795 x 521 mm (d +56mm door handle)
Voltage Electrical load 230 V, 50/60 Hz approx. 1520 W
Voltage Electrical load 115 V, 50/60 Hz approx. 1520 W
Ambient temperature 10 °C to 30 °C
Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Humidity rh max. 70 %, non-condensing
Altitude of installation max. 2,000 m above sea level
Overvoltage category II
Pollution degree 2

Product Support Documents

Find essential brochures, technical data, and user manuals for this product, providing you with quick, reliable, and detailed information to support your needs.

pdf

Memmert Product range brochure

Size: 2.62 MB
Date modified: 29-05-2024
pdf

Memmert Humidity chamber HCP50 data sheet

Size: 1.59 MB
Date modified: 29-05-2024
pdf

Memmert Climate chambers brochure

Size: 9.44 MB
Date modified: 29-05-2024

We are here to help.

With our expert team, cutting-edge technology, unwavering dedication and comprehensive sales portfolio, we ensure your laboratory equipment always operates at peak performance. Whether you're seeking routine maintenance, ad-hoc repairs, precision calibration or the latest in laboratory technology, we're your trusted partner.

This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies.

Laboratory equipment maintenance, repair, calibration and sales.

Request a quote

Your details
Quote details
Explore
Drag