Skip links

Memmert ICH110L

The Memmert ICH110L climate chamber ensures precise environmental control from 0 to +60°C, with active humidity control from 10-80% rh and integrated light for versatile applications. TwinDISPLAY provides advanced control and monitoring for reliable

020 8663 4610
Sales support

Product details

The Memmert ICH110L climate chamber combines precise temperature and humidity control with advanced lighting options, making it ideal for stability testing and research. Key benefits include:

Technical data

Specification Details
Working temperature range without humidity: 0 to +60 °C
Setting temperature range 0 to +60 °C
Working temperature range from +10°C up to +60°C with humidity and/or light
Setting accuracy temperature 0.1 °C
Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity humidity supply with distilled water from external tank by self-priming pump
Humidification humidification by hot steam generator
Humidity active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity – resolution of display 0.1 %, setting accuracy 0.5 %
Light illumination unit consisting of 3 fluorescent lights with cold white light (standard illuminant D65, 6.500K) and 2 UV-lamps in the spectral range of 320-400 nm
Light daylight and UV-light can be programmed individually via controller
Light cold white light +/- 6 %, for UV-light +/- 10 %
ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting German, English, Spanish, French, Polish, Czech, Hungarian
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT the process time does not start until the set temperature is reached
Adjustable parameters temperature (Celsius or Fahrenheit), fan speed, light, programme time, time zones, summertime/wintertime
Fan forced air circulation in working chamber adjustable from 10 to 100 % in 10 % steps
Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
Temperature control over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY additionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Autodiagnostic system integral fault diagnostics for temperature and humidity control
Alarm visual and acoustic
Air jacket air jacket heating system for gentle all-around heating
Heating concept no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior
Defrosting highly efficient automatic defrosting system
Works calibration certificate standard values +10 °C and +37 °C, 60 % rh at +30 °C
Entry port Entry port (silicone), 40 mm clear diameter, moisture tight, can be closed by a silicone stopper, standard position at the back
Internals 2 stainless steel grid(s), electropolished
Standard accessories Water tank including connection hose
Stainless steel interior Dimensions w(A) x h(B) x d(C): 560 x 480 x 400 mm (d less 33 mm for fan)
Volume 108 l
Max. number of internals 5
Max. loading of chamber 150 kg
Max. loading per internal 20 kg
Textured stainless steel casing Dimensions w(D) x h(E) x d(F): 745 x 1233 x 585 mm (d +56mm door handle)
Installation on lockable castors
Housing rear zinc-plated steel
Voltage Electrical load 230 V, 50/60 Hz approx. 1450 W
Voltage Electrical load 115 V, 50/60 Hz approx. 1450 W
Ambient temperature 15 °C to 28 °C (up to 34 °C with limited temperature range)
Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Humidity rh max. 70 %, non-condensing
Altitude of installation max. 2,000 m above sea level
Overvoltage category II
Pollution degree 2

Product Support Documents

Find essential brochures, technical data, and user manuals for this product, providing you with quick, reliable, and detailed information to support your needs.

pdf

Memmert Product range brochure

Size: 2.62 MB
Date modified: 29-05-2024
pdf

Memmert Climate chambers brochure

Size: 9.44 MB
Date modified: 29-05-2024
pdf

Memmert Climate chamber ICH110L data sheet

Size: 1.44 MB
Date modified: 29-05-2024

We are here to help.

With our expert team, cutting-edge technology, unwavering dedication and comprehensive sales portfolio, we ensure your laboratory equipment always operates at peak performance. Whether you're seeking routine maintenance, ad-hoc repairs, precision calibration or the latest in laboratory technology, we're your trusted partner.

This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies.

Laboratory equipment maintenance, repair, calibration and sales.

Request a quote

Your details
Quote details
Explore
Drag