Setting temperature range | 0 to 70 °C |
Working temperature range | with light: 10 to 40 °C |
Working temperature range | without light: from 0 (at least 20 below ambient temperature) to 70°C |
Setting accuracy temperature | 0.1 °C |
Temperature sensor | 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
ControlCOCKPIT | TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function SetpointWAIT | the process time does not start until the set temperature is reached |
Calibration | three freely selectable temperature values |
adjustable parameters | temperature (Celsius or Fahrenheit), programme time, time zones, summertime/wintertime |
Convection | forced ventilation by Peltier fan |
Documentation | programme stored in case of power failure |
Programming | AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Temperature control | over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display |
AutoSAFETY | additionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature |
Autodiagnostic system | for fault analysis |
Alarm | visual and acoustic |
Peltier | Heating and cooling performance distribution by individual control of the Peltier elements in the upper and lower row |
Peltier | energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle) |
Works calibration certificate | for 10 °C, 25 °C and 40 °C |
Door | fully insulated stainless steel door with 2-point locking (compression door lock) |
Door | inner glass door |
Internals | 2 stainless steel grid(s), electropolished |
Dimensions | w(A) x h(B) x d(C): 560 x 480 x 400 mm (d less 32 mm for fan – Peltier) |
Volume | 108 l |
Max. number of internals | 5 |
Max. loading of chamber | 150 kg |
Max. loading per internal | 20 kg |
Dimensions | w(D) x h(E) x d(F): 745 x 864 x 555 mm (d 56mm door handle & 111mm Peltier element) |
Housing | rear zinc-plated steel |
Voltage Electrical load | 230 V, 50/60 Hz |
Voltage Electrical load | 115 V, 50/60 Hz |
Set Up | The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. |
Ambient temperature | 16 ºC to 40 ºC |
Humidity rh | max. 70 %, non-condensing |
Altitude of installation | max. 2,000 m above sea level |
Overvoltage category | II |
Pollution degree | 2 |